High-Frequency Electrosurgical Unit HFE-300 "Nadiya-4"
The device is designed for cutting and coagulation of soft tissues and vessels using high-frequency current without anesthetics during surgical procedures in hospitals and clinics.
The unit is intended for use in specially equipped rooms at ambient temperatures from 10°C to 35°C and relative humidity up to 80% at 25°C.
Operating Principle
The device serves as a source of high-frequency current. HF voltage is applied to the active and passive (monopolar) electrodes or to the bipolar forceps (electrode) and is delivered to the patient’s tissue. The electrosurgical effect of cutting and coagulation is based on generating sufficient thermal energy in the tissue via a narrow HF current stream at the contact point of the active electrode or between the tips of the bipolar forceps. In monopolar mode (CUT-1, CUT-2, COAG-M), the active electrode is significantly smaller than the passive one, resulting in higher current density and localized heating near the active electrode. At sufficient power, tissue fluid turns into steam, causing cellular rupture — this is the cutting effect. In bipolar mode (COAG-B), both electrodes are active, of equal size, and placed close together. HF voltage applied between them heats the tissue. At a certain temperature, tissue proteins coagulate, achieving hemostasis.
Technical Specifications and Additional Features
The device provides the following modes: standby, cutting, and coagulation, with specific operation modes for:
Cutting:
- CUT-1 mode – monopolar, continuous HF oscillation after generator activation;
- CUT-2 mode – monopolar, amplitude-pulse modulation of HF oscillation after generator activation.
Coagulation:
- COAG-M mode – monopolar, continuous HF oscillation after generator activation;
- COAG-B mode – bipolar, continuous HF oscillation after generator activation.
Output power is adjustable in all modes — from minimum to maximum.
The device provides:
- Idle and short-circuit modes at any output power level;
- Light indicators:
- Power supply indicator – green;
- Selected operation modes: CUT – yellow on a yellow field; COAG – yellow on a blue field;
- Output power level – red indicator on yellow or blue fields depending on mode;
- Passive electrode cable disconnection – red "Err PE" warning on power level indicators.
- Audible alerts:
- Continuous sound during CUT mode power output; intermittent tone during COAG mode;
- Double intermittent tone if the passive electrode cable is disconnected.
- The unit operates at maximum output power for 8 hours under optimal load in intermittent duty cycle: 10 seconds of HF current output, followed by a 30-second pause;
- Maximum continuous connection to power grid – no more than 8 hours.
The device is protected against defibrillator discharge.
The device package contents are presented in the table below
| No. | Item | Designation | Quantity |
| 1 | Electronic unit | 1 | |
| 2 | Control pedal (dual-button) | B-001 | 1 |
| 3 | Passive (neutral) electrode | B-003 | 1 |
| 4 | Monopolar electrode holder (2.0 mm) | B-005 | 1 |
| 5 | Bipolar forceps holder | B-007 | 1 |
| 6 | Bipolar forceps, straight, 200 mm | PP-007 | 1 |
| 7 | Straight knife electrode | 04500 | 1 |
| 8 | Curved knife electrode | 04500 | 1 |
| 10 | Device passport | 1 | |
| 11 | Device packaging | 1 |
Device Design
The unit consists of two separate components: the electronic block and the footswitch.
Front Panel of the Electronic Unit |
![]() |
1. Manufacturer's trademark
2. Model name
3. CUT-2 mode indicator – mixed cutting
4. CUT-1 mode indicator – pure cutting
5. Digital output power display for CUT mode
6. COAG-B mode indicator – bipolar coagulation
7. COAG-M mode indicator – monopolar coagulation
8. Digital output power display for COAG mode
9. Button to increase power level in COAG mode
10. Button to decrease power level in COAG mode
11. Connector for bipolar forceps cable
12. COAG mode selector button
13. Button to increase power level in CUT mode
14. Connector for monopolar active electrode cable
15. Button to decrease power level in CUT mode
16. Working part identification symbol
17. Connector for passive electrode cable
18. CUT mode selector button
19. BF type and defibrillator protection symbol
20. Power switch
Rear Panel of the Electronic Unit |
![]() |
21. Connector for footswitch
22. Class II electrical safety symbol
23. Power supply cable
The electronic unit is enclosed in a plastic case consisting of upper and lower covers and front and rear panels. On the front panel, the following controls and indicators are located:
- Power toggle switch and "POWER" indicator (20);
- Mode selection buttons (12, 18) and their indicators (3, 4, 6, 7);
- Power adjustment buttons (9, 10, 13, 15) and digital displays (5, 8);
- Connector "PAS" (17) for passive electrode cable;
- Connector "ACT" (14) for monopolar active electrode holder;
- Connector "BI" (11) for bipolar forceps holder.
On the rear panel, the power cable (23) and "PEDAL" connector (21) for the footswitch are located.
Footswitch |
|
![]() |
|
|
a) Top view 1. Cable connecting the pedal to the electronic unit 2. Pedal for activating CUT mode 3. Pedal for activating COAGULATION mode |
b) Bottom view 4. IPX7 protection code symbol |
The footswitch is a floor-standing construction with two movable platforms: the left one (2) activates the CUT mode, and the right one (3) activates the COAGULATION mode.
The platforms are spring-loaded and move vertically: downward (to activate the mode) when pressed with the foot, and upward (to deactivate the mode) when released. The footswitch connects to the electronic unit via a cable.
High-Frequency Electrosurgical Unit HFE-300 “Nadiya-4”: Precision and Safety Guaranteed
This device is widely used in veterinary surgery. The compact unit is suitable for a variety of procedures, including cutting and coagulation. You can order the HFE-300 electrosurgical unit “Nadiya-4” online from our website, complete with manufacturer warranty and certified performance.

How the HFE-300 “Nadiya-4” Electrosurgical Unit from Our Online Catalog Works
This high-frequency device is commonly used in veterinary surgical clinics. It’s ideal for open surgeries, laparoscopic procedures, and other surgical applications. You can order the HFE-300 “Nadiya-4” electrosurgical unit for veterinary purposes directly from our store — at competitive prices.
How does thermal impact occur during surgery?
When the active electrode contacts the patient’s body, high-frequency current heats up the cells. This causes intracellular fluids to boil and evaporate instantly, while proteins coagulate and vessel walls fuse. This results in effective hemostasis — bleeding is stopped even when vessels are incised.
In professional veterinary practice, purchasing an HFE-300 “Nadiya-4” electrosurgical unit in Ukraine is considered one of the best options for safe surgical operations. Its generator produces high-frequency current — relevant for over 80% of veterinary procedures today. The unit can be used for:
- Coagulation – gentle sealing of soft tissues and blood vessels. The surface is heated to 60–100°C, causing both intracellular and extracellular fluids to evaporate. The cell structure remains intact, which is essential for preserving hemostasis.
- Tissue cutting with high precision and minimal trauma or bleeding. Muscular and adipose tissues can be incised using voltage starting from 200V and power over 8W, utilizing electric arc technology. Electrodes can be shaped as loops, needles, or spatulas.
- Electroligature – restoring vessel integrity via thermal sealing. You can order the HFE-300 “Nadiya-4” electrosurgical unit online to safely seal small vessels. For optimal precision, forceps-shaped electrodes are recommended.
- Devitalization – removal of pathological tissues and organs through point-specific destruction. Affected tissues gradually degrade and are resorbed via metabolic processes after exposure to high-frequency electricity.